Specifications
|
|
Functional Specifications
|
|
Transducer Calibration Accuracy
|
Available to ±0.1kg/m3 (±0.0062 lb./ ft.3)
|
|
Repeatability
|
0.02kg/m3 (0.0012 lb./ ft.3)
|
|
Flow Range
|
Vertical installation: 0L/min to 300L/min (0 USG/min to 79 USG/min); Horizontal installation: 5L/min to 300L/min (1.3 USG/min to 79 USG/min)
|
|
Operating Density Range
|
0 kg/m3 to 2100kg/m3 (0 lb. / ft.3 to 131.1 lb. / ft.3)
|
|
Installation
|
Vertical installation (standard), horizontal installation (optional); No instrument or pipework supports required
|
|
Pressure Effect (corrected)
|
0.003 kg/m3/bar (0.000013 lb. / ft.3/psi) note: correction coefficients applied
|
|
Temperature Effect (corrected)
|
0.005kg/m3/°C (0.0002 lb./ ft.3/°F) note: correction coefficients applied
|
|
Density Meter Dimensions
|
Dia. x L: 3.5 x 20.87 in. (89 x 530mm)
|
|
Shipping Dimensions
|
24 x 16 x 12 in. (approx 590 x 390 x 290mm)
|
|
Net Weight
|
24 lb. (11kg)
|
|
Shipping Weight
|
33 lb. (15kg)
|
|
Environmental Rating
|
IP65 (NEMA 4X)
|
|
Electrical Connections
|
Screw terminals; Cable entry: 2 x 0.75 in. NPT
|
|
Temperature Measurement
|
High accuracy 1⁄3 DIN integral 4-wire PT100
|
|
Local Display (H option)
|
41⁄2-digit 0.3 in. (7.6mm) 7-segment LCD display. Resolution 0.1% or 0.01% depending on display variable.
|
|
Secondary Containment
|
As flange rating to Class 300 then 2.5 times maximum safety flange rating to Class 600
|
|
Factory Calibration Range
|
40.58 lb. / ft.3to 99.98 lb. / ft.3 (650kg/m3 to 1600kg/m3)
|
|
Ambient Temperature Range
|
-20°C to +60°C (-4°F to +140°F) ambient
|
|
Process Temperature Range
|
Sarasota FD910 / FD950 / FD960: -50°C to +180°C (-58°F to +356°F)
|
|
Output
|
F option (frequency output): Frequency related to density on 2-wire current modulated loop 6 mA to 18 mA; 4-wire PT100; H option (headmounted electronics): Analog 4-20 mA related to density or density derived variable; HART protocol
|
|
Power Supply
|
F option (frequency output): 13-28VDC 10mA average (peak 18mA); H option (headmounted electronics) 2 x 13-28VDC 25mA; 4-20mA current pressure input available
|
|
Maximum Operating Pressure
|
Sarasota FD910 / FD950 / FD960: as flange rating
|
|
Material Specifications
|
|
Sensor[s]
|
Sarasota FD910 stainless steel (316L / 1.4404); Sarasota FD950: Hastelloy C276; Sarasota FD960: Ni-Span C
|
|
Other Wetted Parts
|
Sarasota FD910 (Class 150, 300) / FD960: 316L stainless steel (316L / 1.4404)
Sarasota FD910 (Class 600) / FD950: Hastelloy C276
|
|
Case
|
Stainless steel (316L / 1.4404)
|
|
Electronics Housing
|
Copper free aluminium grey epoxy finish; Plate glass window for local display option
|
|
Process Connections
|
1-in ASME B16.5 RF (raised face): Sarasota FD910 / FD960: stainless steel (316L / 1.4404) – Class 150, 300 or 600; Sarasota FD910 / FD960: duplex (A 182 Gr.F51) – Class 150, 300 or 600; Sarasota FD950 only: Hastelloy C276 – Class 150, 300 or 600
25mm BSEN1092 RF (raised face - type B): Sarasota FD910 / FD950 / FD960: Up to maximum PN100
Other Flange Types: Consult Thermo Fisher Representative
|