Specifications
|
|
Performance Specifications
|
|
Precision
|
From ±0.001gm/cc depending upon application
|
|
Repeatability
|
±0.001gm/cc depending upon application
|
|
Stability
|
Drift less than ±0.05% of radiation change over 6 months
|
|
Resolution
|
Analog output 16-bit Internal instrumental IEEE754 floating point
|
|
Magnetic Field Effect
|
No effect on output when tested to IEC 770 para. 6.2.11
|
|
Ambient Temperature Field
|
±0.09% of radiation change per °C; ±0.06% of radiation change per °C
|
|
Surface Radiation
|
Dependent on the application, source selection and physical installation
|
|
Power
|
115/230VAC ±10%; 50 to 60Hz
|
|
Power Supply Effect
|
No effect on operation over specified power supply ranges
|
|
Gamma Ray Source
|
|
Type
|
Cs-137 or Co-60, both stainless steel, doubly encapsulated
|
|
Size
|
10 to 10,000mCi Cs-137 or 1000 to 3000mCi Co-60
|
|
Functional Specifications
|
|
Detector
|
PVT plastic scintillator with wide dynamic range (PVT resists shock and moisture damage); Stabilization via electronic control without heater stabilization for optimum performance
|
|
Operating Temperature Limits
|
Transmitter/Detector: -40° to +60°C (-40° to +140°F) ambient
|
|
Storage Temperature Limits
|
-30° to +65°C (-22° to +149°F)
|
|
Inputs
|
Flowmeter: 4 to 20mA linear; Dry contact closure; Temperature compensation circuitry with 100ohm platinum RTD 2 or 3 wire
|
|
Current Outputs
|
4 to 20mA isolated, self-powered into 700ohms, field scalable RS 485 duplex
|
|
Serial Outputs
|
RS 423 or Bell 202 interface with HART* protocol; Digital signal superimposed on 4 to20mA signal using HART protocol, available for control system interface; Digital signal available via intrinsically safe connection for HART Communicator
|
|
Contact Closure Inputs
|
Two programmable dry contact inputs
|
|
Contact Closure Outputs
|
Two 115VAC/28VDC SPDT at 10A (230VAC SPDT at 8A)
|
|
Source Decay Compensation
|
Selectable Cs-137, Co-60 or Am-241
|
|
Linearization
|
Logarithmic function and ratio linearization
|
|
Analog Output Adjustment
|
Engineering units and range values user-selected; All analog trim and adjustments made through software commands
|
|
Adjustable Damping
|
1 to 1040 sec.
|
|
Adaptive Damping
|
See Dynamic Process Tracking
|
|
Dynamic Process Tracking
|
Selectable; Automatic 8 to 1 time constant reduction based on process rate of change
|
|
Transmitter
|
System architecture: Multiprocessor-based electronics provide uninterrupted output during data entry and system interrogation; Surface-mounted technology provides high degree of reliability; All user data doubly stored in non-volatile memory with no battery backup required; Programming options: Menu-driven offers direct keypad entry; HART Smart Model 375 handheld terminal: used to setup and calibrate gauge, and to enter data; Communicates with any
DensityPRO via the current loop; BEL202FSK standard; Thermo Scientific 9734 handheld terminal: used to setup and calibrate gauge, and to enter data; communicates with any DensityPRO via the RS485 connector; provides upload/download of gauge configuration to/from PC via RS232 interface; Comm PC Interface Software: interface with up to 32 DensityPRO units over RS485 loop (RS485 to RS232 converter provided)
|
|
Physical Specifications
|
|
Electrical Connections
|
Two 0.75 in. NPT fittings; HART Communicator interface connects directly into 4 to 20mA loop
|
|
Interface Cable
|
Shielded 14 to 22 gauge Belden 9318
|
|
Transmitter Display
|
Thermo Scientific 9734 handheld terminal
|
|
Gauge Mounting Hardware
|
Gamma ray source: integral bolt-on bracket; Compatible with chain or saddle mount; Transmitter/Detector: single or dual chain universal mount 63.5mm (2.5 in.) to 914.4mm (36 in.); Pipe saddle mount 50.8mm (2 in.) to 1066.8mm (42 in.); Tabs for mounting on insulated pipes
|
|
Housing
|
Detector: Carbon steel polyurethane painted or optional SS 316; Source: carbon steel or optional SS 316, lead filled, polyurethane painted; two position shutter, locks in closed position
|
|
Weight, Detector/Electronics
|
4.5kg (10 lb.)
|