Thermox analyzers from Ametek are among the most established combustion monitoring and control devices available today. A broad range of products is available for analyzing oxygen, combustibles, and methane in combustion flue gases for combustion control, safety, and thermal NOx reduction. Industries served include Electric Power Generation, Petroleum Refining, Petrochemical Processing, Cement & Lime Manufacturing, Glass and Ceramics, Pulp
& Paper, Iron & Steel, and Non-Ferrous Metals production.
|
|
Analyzers for Continuous Emissions (Pollution) Monitoring Systems and Wet/Dry ESP outlet flue gas scrubbing. Component Wet/O2 Analyzers for CEM Systems Integration
CEM/O2 Download printable brochures: CEM O2 - CEM Humox - RM CEM
O2 - CEM TM
The Need
Continuous Emissions Monitoring systems using hot-wet extractive systems take the flue gas sample from the exit of wet-scrubbers. A hot-wet oxygen measurement is taken from the incoming feed line to the CEM system analysis cabinet. The oxygen analyzer used for this measurement must be equipped to keep the sample gas above the dew point and compact to fit within the restricted space of the analysis cabinet.
The Solution
The CEM O2 sensor was designed to meet the replacement and integration needs of hot-wet extractive CEM systems integrators. The sample inlet feeds a straight-run sample line with center mounted chimney-effect convection stack for continuous sampling. Compact sensor housing fits easily into CEM cabinet.
There are two control options available for the CEM/O2 analyzer.
-
IQ Series Transmitter provides sensor head mounted full function electronics with analog outputs, alarms, expanded system diagnostics, auto calibration feature, RS485 and optional Hart communication.
-
2000 Series Microprocessor Control provides 4 line local display, analog outputs, alarms, expanded system diagnostics, auto calibration option and RS485 communication, and optional Humox O2 wet/dry calculation software for correlated flue gas moisture measurement.
|
|
|
|
Specifications
|
|
Diagnostics:
|
Watchdog Timer
Service Alarm (trouble alarm)
|
|
RS-485 (Analyzer) Port
|
Supports up to 16 devices
|
|
Distance:
|
4000 feet (1200 m) max.
|
|
Media:
|
single shielded, twisted-pair,18-24 AWG
|
|
Network Ports
|
|
|
Ethernet:
|
IEEE 802.3 10/100 Base-T, auto-sensing
|
|
Media:
|
Category 5e UTP or STP (recommended)
|
|
|
Foundation Fieldbus H1
|
|
Protocols Supported:
|
HTTP (web pages)
FTP (software updates)
SMTP (e-mail)
POP3 (e-mail)
|
|
Modbus
|
|
Foundation Fieldbus
|
|
Browsers Supported:
|
Microsoft Internet Explorer V5.0 or higher, Netscape Communicator V4.73 or higher
|
|
Power Requirements:
|
100-240 VAC, 50/60 Hz
Alarm Contacts (4) 1 Amp at 24 VDC or .5 Amp at 125 VAC, noninductive load (1) contact configured for system alarm
|
|
Environment
|
|
Ambient
Temperature:
|
32°F to 122°F (0°C to 50°C)
|
|
Relative Humidity:
|
10% to 80%, non-condensing
|
|
IEC Installation:
|
Category II
|
|
IEC Pollution:
|
Degree 2
|
|
Enclosure:
|
Indoor/outdoor use UL Type 4X and IP65 Hazloc. This device is for use only in General Purpose Areas
|
|
Display Information for Each Analyzer
|
Gas concentration for each measured species Alarm status, Principal diagnostic parameters, Actual parameters vary depending on the capabilities of the supported analyzers
|
|
Security:
|
Password protection to prevent unauthorized access to parameters and settings.
|
|
|
|
Ametek Main Page - Thermox - Western Research - Process Instruments - Dycor - Land - Chandler - Trace Analytical
|
|
|