Specifications
|
|
Measurement Ranges:
|
|
CO Ranges:
|
0 - 50 ppm up to 0 - 2 000 ppm
|
|
Resolution:
|
1 ppm / 1 mg/m3
|
|
Linearity:
|
2% of range
|
|
Zero drift:
|
2% of range per month
|
|
Span drift:
|
2% of range per month
|
|
02 Ranges:
|
0 - 5 % to 0 - 25%
|
|
Resolution:
|
0.1 Vol %
|
|
Linearity:
|
0.2 Vol %
|
|
Zero drift:
|
0.2 Vol % per month
|
|
Span drift:
|
0.2 Vol % per month
|
|
Calibration:
|
Microprocessor controlled
|
|
Standard:
|
Manual gas selection
|
|
Display:
|
LCD with backlight
|
|
Indicators
|
External "power On" and "System OK" LEDs
|
|
Outputs/Inputs
|
|
Analog outputs:
|
Isolated current loop outputs
Efficiency (on FGA 930E only)
0, 2, or 4 to 10 or 20 mA
|
|
Relay outputs:
|
System OK; Maintenance/Calibration
|
|
Relay rating:
|
Isolated changeover S.P - 1 A @ 240 V a.c.
5 A @ 240 V d.c. resistive for Alarm - one per gas measured
Span and Zero Gas Relays for calibration gases
|
|
Inputs:
|
Current loop inputs for ambient and process temperatures (only needed for FGA 930 E)
|
|
Environmental
|
|
Environmental rating:
|
IP65 / NEMA 4
|
|
Operating (ambient) temp:
|
32 to 95ºF / 0 to +35ºC standard to -4ºF / -20ºC with optional case heater to 122ºF / +50ºC with optional air conditioner
|
|
Compliance
|
|
Measurement standards
|
Meets the requirements of ISO 12039, ISO 10849 and ISO 7935
|
|
Approvals:
|
MCerts for Continuous Emissions Monitoring Systems (Nov 1988)
TUV approved to the 13th & 17th BlmSchV and TA Luft USEPA certified to PS-3, PS-4 and PS-4A
|
|
Electrical Safety:
|
Conforms to EN-61010-2
|
|
EMC:
|
Conforms to EN-50 081 & EN-50 082
|
|
Power
|
|
Power Supply:
|
110 V a.c. or 230 V a.c., +20 %, 50 - 60 Hz
|
|
Power consumption:
|
300 W
|
|
Gas and Air Requirements
|
|
Instrument air (zero calibration):
|
2 bar / 30 psi clean and dry, 5 l/min / 0.2 cfm
|
|
Instrument air (cooling):
|
2 - 10 bar / 30 - 150 psi clean and dry, 90 l/min / 3 cfm
|
|
Calibration gas (recommended):
|
2 bar / 30 psi, 5 l/min / 0.2 cfm 20 litres / 0.7 cu. ft. per calibration approx.
|
|
Calibration gas type:
|
Specific to each gas type and measurement span
|
|
Dimensions (HxWxD):
|
24 x 24 x 14 in / 800 x 350 mm
|
|
Weight:
|
117 lb / 53 kg
|
|
Options
|
Special measurement ranges
Automatic calibration
Efficiency measurement (FGA 930 E only)
Probe type
Sample line type
Case heater
Air conditioner/vortex cooler
|