ETA Process Instrumentation - since 1971

ETA Process Instrumentation -- Since 1971

ETA Associates merged with Process Instrumentation Inc. as of 9/1/2011. Visit our sister companies: Safety Inc and iFacility Service.

     Home

  Manufacturers 

Industries & Instruments

Technical Resources

    Service   

     About      

Ametek Thermox Ametek Chandler Engineering  
 

Chandler Engineering includes dew point testers, moisture analyzers, natural gas chromatographs, liquid densitometers, and gas gravitometers.

Ametek Chandler Engineering UGC Densitometer

UGC Vibrating Densitometer                                 Download printable brochure.

The UGC Vibrating Densitometer accurately measures the density of a flowing fluid at actual operating conditions. This Densitometer was primarily developed for extremely accurate fluid metering (custody transfer), plus many other applications, such as pipeline interface detection, blending, process control, etc. The frequency output of the Vibrating Densitometer is compatible with most flow computers.


Specifications
Accuracy: 1. +/- .0002 gr/cc max. from .3 to .5 gr/cc; +/- .0125 from 18.7 to 31 lb/ft3
2. +/- .0004 gr/cc max. from .5 to 1.0 gr/cc; +/- .025 from 31 to 62 lb/ft3
3. +/- .0005 gr/cc max. from 1.0 to 1.6 gr/cc; +/- .031 from 62 to 100 lb/ft3
Repeatability: 1. +/- .0001 gr/cc max. for .3 to 1.0 gr/cc density range; +/- .0062 from 18.7 to 62 lb/ft3
2. +/- .0002 max. for 1.0 to 1.6 gr/cc density range; +/- .0125 from 62 to 100 lb/ft3
Calibrated Density Range: .3 - 1.0 gr/cc (300 to 1000 kg/m3)(18.7 to 62.43 lb/ft3)
  Optional: To 1.6 gr/cc (1600 kg/m3)(100 lb/ft3)
Temperature Range: -40 to 85°C (-40 to 185°F)
Process Connections: 3/4” FNPT. Specify pipeline size for optional scoops.
Recommended Flow Rate: Minimum 5.7 LPM (1.5 GPM)
Maximum Working Pressure: FP27800 (Ni-Span C)
4000 PSI (276 BARS)
FP27800 (SS)
3200 PSI (220 BARS)
FP27800 (Hastelloy)
3200 PSI (220 BARS)
Power Requirements: 15 to 36 VDC (24 VDC ± 8 VDC @ 35 mA)
Output Signal: The open collector output (referenced to power supply common) is capable of sinking up to one Amp of current (35 mA is nominal). Typical frequency range from 1 KHz - 2 KHz. Pull-up resistor is required when used with typical flow computers.
Size: 100mm round, 460mm long (4” x 18 1/8”)
Manifold: Model 278: 203mm x 50mm x 50mm (8” x 2” x 2”)
Shipping Weight: 25 lb (11 kg)
Wetted Parts: Ni-Span C Tubes; 304 stainless steel manifold; Viton O-rings.
Standard Cover: NEMA 4, alodine-coated and powder-coated aluminum.

Back to Chandler Engineering - Ametek Main Page -  Thermox - Western Research - Process InstrumentsLand - Trace Analytical

Ametek Badger Meter Baseline Mocon Berkeley Nucleonics • BH Thermal Corp Brandt - Thermo Bud Industries Ceasefire Chandler Engineering - Ametek • Daniel Industries Dent Instruments Draeger Gas Detection Systems • Dycor - Ametek • FasTest • Fox Thermal Instruments Gasmet H2Scan Hawk Measurement Systems Horiba ION Science JMS Southeast • Kahn Instruments Kanomax Lakewood Instruments Land - Ametek • Magnatrol Valve Martel Electronics/Beta Calibrators • Measurement Technologies Odor Eyes • Percival ScientificPerma Pure LLCProcess Instruments - Ametek • Process Sensors • Racine/Preso Flow MeteringR. StahlRKI InstrumentsRosemount Tank GaugingSarasota - Thermo • SpectiteTexas Nuclear Kay/Ray Sensall - Thermo TestoThermo • Thermox - Ametek • Trace Analytical - Ametek United Electric ControlsWelkerWestern Research - Ametek • Westronics - Thermo • Wilkerson Instrument Co 

ETA Process Instrumentation, 119 Foster Street, Bldg. 6, Peabody, MA 01960 • (978) 532-1330 or sales(a)etapii.com