Chandler Engineering includes dew point testers, moisture analyzers, natural gas chromatographs, liquid densitometers, and gas gravitometers.
|
|
Model 292 Portable BTU Analyzer Chromatograph Download printable brochure.
The Need
Natural gas quality is an increasingly significant issue for both pipeline operations and integrity. Operators need a reliable, easy to use means to verify gas quality, or the performance of on-line gas quality analyzers, at remote locations. A portable, reliable, and accurate means of gas quality analysis is needed.
The Solution
Chandler has eliminated the hassle and inaccuracy of portable gas analysis. Now field technicians can perform the required gas quality tests with ease where they are most accurate ... on site. On-site analysis not only eliminates the concerns over accurate sample acquisition and transport, it also eliminates the DOT concerns over transporting natural gas samples.
The 292 is designed around field-proven chromatographic techniques that are defined by ASTM, GPA, AGA, and ISO. This portable BTU Analyzer Chromatograph is truly "operator friendly." Sophisticated software does all the work of compositional analysis, and calculations of BTU value, Wobbe index, compressibility, and relative density (specific gravity). Results are available on the panel digital display and can be printed or transferred to a computer. The
analysis is very easy to conduct with a simple sample line connection and a press of the START TEST button.
The 292's internal data storage can retain information about multiple analyses from dozens of different test locations.
|
|
|
|
Specifications
|
|
Performance
|
|
Range:
|
0-2,000 BTU/scf and 0- 74,500 kJ/m3
|
|
Repeatability:
|
±0.05% or ±0.5 BTU/1000 BTU
|
|
Standards:
|
Meets applicable GPA, ASTM, AGA-8 (direct method) and ISO standards.
|
|
Approved for:
|
Custody Transfer by Industry Canada, California Public Utility Commission and GOSSTANDART of Russia.
|
|
Environmental
|
|
Operating Temp:
|
-10° to 130°F (-25-55°C), non-condensing
|
|
Enclosure:
|
NEMA 4
|
|
Power Requirements:
|
115 or 230 VAC ± 10%, 50/60 Hz 250 VA startup, 40 VA operating
Can be operated with inverter from 12 VDC
|
|
Sample Gas:
|
Analyzes liquid-free natural gas that contains nitrogen, methane, carbon dioxide, ethane, propane, I-butane, N-butane, I-pentane, neopentane, N-pentane, C6 and heavier components.
|
|
Max Inlet Pressure:
|
3,500 psig (24,132 kPa)
|
|
Min Inlet Pressure:
|
7-10 psig (48-69 kPa)
|
|
Connection:
|
1/8” stainless steel tube
|
|
Carrier Gas:
|
Zero-grade helium (99.995% pure)
|
|
Pressure:
|
150 psig (1135 kPa)
|
|
Connection:
|
1/8” SS tube (metric adapter kit available)
|
|
Supply Gas (GC valve):
|
Nitrogen, helium, or instrument air
|
|
Pressure:
|
80 psig (552 kPa)
|
|
Connection:
|
1/4” flared tubing
|
|
Calibration Gas:
|
Certified natural gas standard in 14-liter bottle (240 psig) consisting of 11 components: approximately 90% methane (1050 BTU value). Special handling for hazardous material is required for shipment. Part #292-00342
|
|
Analog Output Signals:
|
|
Chromatogram:
|
0-10 VDC, 1000 Ohm impedance
|
|
Oven temperature:
|
0-10 VDC, 1000 Ohm impedance
|
|
4-20 mA Outputs:
|
Model 298 transmitter option provides four 4-20 mA outputs per stream. Connects to serial port.
|
|
Printer Interface
|
|
Parallel Port:
|
Epson-compatible interface to dot-matrix, laserjet or inkjet printer.
|
|
Serial Port:
|
RS-232C, 300-9600 baud, non/even/odd parity, 7 or 8 word length, 1 or 2 stop bits, modem compatible, DTE.
|
|
|
Back to Chandler Engineering - Ametek Main Page - Thermox - Western Research - Process Instruments - Land - Trace Analytical
|
|
|