Baseline MOCON, Inc. markets a full line of gas detection equipment including the award winning piD-TECH® Plug-In Photoionization Sensor, Model 9000 and Model 8800 Total Hydrocarbon/VOC Analyzers, and Model 8900 automated online Gas Chromatographs 

 

 Baseline 9000 Total Hydrocarbon Analyzer

Series 9000 Total Hydrocarbon Analyzer

Download printable brochures:  Total Hydrocarbon Analyzer - Non Methane Hydrocarbon Analyzer - TCA Total Hydrocarbon Analyzer - Heated Hydrocarbon Analyzer

Overview

The Baseline® Series 9000 is the candidate of choice whenever accurate, reliable total hydrocarbon analysis is required. The Series 9000 analyzer provides unparalleled flexibility and offers continuous, fully automated gas analysis over a broad range of concentrations.

The 9000 has a dynamic range from less than 10 ppb on the low range version and up to 100% as methane on the high range version. The analyzer has a generous complement of analog, digital and logic output capabilities.

The analyzer is based on a flame ionization detector (FID) with FlowGuard electronic control that delivers a small portion of the sample gas to the detector flame. During the combustion process, organic or hydrocarbon-based gases in the sample are ionized then they are detected by the instrument and reported as a concentration.

The Series 9000 can be configured with internal components for single or multipoint analysis of non-condensing gas samples. The automatic calibration feature enhances the long-term analytical stability of the instrument. These features place the instrument well ahead of the competition in performance, automation, and configurability.

 

Applications

The Series 9000 is designed to continuously monitor the total hydrocarbon content of non-condensing gases in a variety of applications, such as:
  • Beverage grade CO2 analysis
  • Scrubber & oxidizer efficiency
  • Carbon bed breakthrough detection
  • Contaminant analysis in pure/ultrapure inert gases
  • Well logging
  • Industrial hygiene & safety monitoring
  • Fenceline (perimeter) monitoring around industrial sites
  • LEL monitoring

Available configured for: NMHC NonMethane Hydrocarbon Analyzer, TCA Total Carbon Analyzer, H Heated Hydrcarbon Analyzer

 

Features

  • Flame ionization detector.
  • 3U case, 19" (48.3cm) rack mountable.
  • Automatic calibration at user-defined intervals.
  • Internal, multipoint sampling option.
  • FlowGuard electronic control of fuel, air and sample.
  • Electronic back-pressure regulator with sample bypass system ensures fast response.
  • Discrete, multilevel concentration & fault alarms.
  • Analog output ranges programmable from 1ppm - 100% full scale.
  • Programmable relays for concentration alarms, events and diagnostics.
  • Automatic FID ignition, with automatic shut-off of sample, fuel and combustion air.
  • Remote operation via RS-232 and Ethernet.
  • Graphical display and easy to use menu system.
  • Optional 9000 Keeper software to monitor and control the instrument remotely. 
 

 Analytical Specifications

Detector

 Flame ionization detector (FID)
Ranges MDQ 0.003%, Full-scale 0.01 - 100% (Methane)
MDQ 0.01ppm, Full-scale 1 - 200ppm (Methane)
MDQ 0.1ppm, Full-scale 1 - 2,000ppm (Methane)
MDQ 0.3ppm, Full-scale 1 - 20,000ppm (Methane)

Repeatability

 ±1% Full-scale response

Drift

 ±1% of Full-Scale over 24 hours

Response Time

 < 5 seconds to 90% of final reading

Alarms

 Multilevel concentration and fault alarms that result in audible and visual alarms. Alarms may also be mapped to relays to control external equipment.

Sampling

 Internal single or multipoint modules for pre-filtered (<0.1 microns), non-condensing samples, with or without sample pump.

Calibration

 Programmable automatic or manual calibration

Support Gas

 Hydrogen 35 cc/min, Air 175 cc/min (typical). Hydrocarbon content must be less than 1 ppm. Fuel blend options available, consult Baseline®.

 

 Physical Specifications

 

Dimensions

 19.00" (48.3 cm) W x 14.25" (36.2 cm) D x 5.25" (13.3 cm) H
 

Weight

 < 20 lb (9.07 kg)
 

Configuration

 Bench-top or rack-mount 19" (48.3cm) panel
 

Connections

 1/4" (6.35 mm) tube fitting connectors
 

Operating Temperature

 32 to 104°F (0 to 40°C)
 

Operating Humidity

 0 - 95% (non-condensing)

 

 Electrical Specifications

 

Power

 90 - 230 VAC, 50/60Hz, 3A
 

Display

 3.4" (8.6cm) x 4.5" (11.4cm) graphical display
 

Relay Outputs

 5 programmable form A relays rated to 3A at 230V AC (optional additional 9 relay output board)
 

Analog Outputs

  1 programmable 0-20mA or 4-20mA isolated output (optional additional 3 analog output board)
 

Digital Outputs

 RS-232, Ethernet

Home Page - Baseline Main Page

Ametek Badger Meter Baseline Mocon Berkeley Nucleonics • BH Thermal Corp Brandt - Thermo Bud Industries Ceasefire Chandler Engineering - Ametek • Daniel Industries Dent Instruments Draeger Gas Detection Systems • Dycor - Ametek • FasTest • Fox Thermal Instruments Gasmet H2Scan Hawk Measurement Systems Horiba ION Science JMS Southeast • Kahn Instruments Kanomax Lakewood Instruments Land - Ametek • Magnatrol Valve Martel Electronics/Beta Calibrators • Measurement Technologies Odor Eyes • Percival ScientificPerma Pure LLCProcess Instruments - Ametek • Process Sensors • Racine/Preso Flow MeteringR. StahlRKI InstrumentsRosemount Tank GaugingSarasota - Thermo • SpectiteTexas Nuclear Kay/Ray Sensall - Thermo TestoThermo • Thermox - Ametek • Trace Analytical - Ametek United Electric ControlsWelkerWestern Research - Ametek • Westronics - Thermo • Wilkerson Instrument Co 